| ARES Device — Audible Notifications | ||
| Audio Notification | Meaning | What Should you Do |
| Unicorder has been turned off. | Unicorder is turned OFF using the ON/OFF switch. | No action needed |
| Unicorder has been turned on. | Unicorder is turned ON using the ON/OFF switch. | You can now test with the unit or press and hold the ON/OFF switch for 3 seconds to turn off. |
| Warning. The Unicorder is charging. | Unicorder is either plugged into the wall charger via USB or is plugged into the PC via USB port without being recognized | Once the green light is no longer flashing, the Unicorder is fully charged (typically 2 hours) |
| Call Tech Support. The Unicorder is not working. Code 1. | Hardware error detected during start-up that makes recording impossible | Contact the office that provided the device. |
| Call Tech Support. Unicorder internal check failed and study cannot be performed. Code 5. | Firmware check failed on study start-up that makes recording impossible. | Initialize the device to clear error |
| Call Tech Support. The Unicorder is not working. Code 6. | The forehead sensor is not sensing enough tension to perform study or the forehead sensor needs to be replaced. | Adjust the Unicorder on the forehead to clear error. If the message persists, please contact the office that provided the device. |
| The Unicorder battery is low. It must be recharged. | Battery has less than 4 hours of charge remaining | Charge the Unicorder via the Wall Charger |
| Call Tech Support. The memory card is full. Code 7. | Memory card is full, recording is not impossible. | Initialize the device to clear error. If the message persists, please contact Support |
| Lie on your back, look at the ceiling and do not move. | Startup command for Initializing the study once the device has been turned on. | Lie on your back as instructed. Will hear additional audio when okay to move around |
| Initialization successful. You can now go to sleep. | Initialization period has finished. | The patient can situate his/her desired posture and proceed to sleep. |
| Tighten the nasal cannula. | Alarm signaling the nasal cannula has fallen off or that there is inadequate airflow (pressure) causing the alarm to sound. | Patient is instructed to inspect the device and tighten the cannula back in place on the device if necessary. |
| The Unicorder has fallen off. | Alarm signaling the device has fallen off the head. | Reposition the Unicorder on head to resume testing. |
| Adjust the Unicorder on your forehead. | Optical Signal is disrupted. | Adjust the Unicorder on the forehead to clear error. If the message persists, please contact the office that provided the device. |