ARES Device — Audible Notifications |
Audio
Notification |
Meaning |
What Should you Do |
Unicorder has
been turned off. |
Unicorder is turned
OFF using the ON/OFF switch. |
No action needed |
Unicorder has
been turned on. |
Unicorder is turned
ON using the ON/OFF switch. |
You
can now test with the unit or press and hold the ON/OFF switch for 3 seconds
to turn off. |
Warning. The
Unicorder is charging. |
Unicorder
is either plugged into the wall charger via USB or is plugged into the PC via
USB port without being recognized |
Once the green light is no longer flashing, the Unicorder is fully charged (typically 2 hours) |
Call Tech Support. The
Unicorder is not working. Code 1. |
Hardware error detected during
start-up that makes recording impossible |
Contact the office that provided the
device. |
Call Tech Support. Unicorder internal check failed and study
cannot be performed. Code 5. |
Firmware
check failed on study start-up that makes recording impossible. |
Initialize
the device to clear error |
Call Tech Support. The
Unicorder is not working. Code 6. |
The forehead
sensor is not sensing enough tension to perform study or the forehead sensor
needs to be replaced. |
Adjust the
Unicorder on the forehead to clear error.
If the message persists, please contact the office that provided the
device. |
The Unicorder battery is low.
It must be recharged. |
Battery has
less than 4 hours of charge remaining |
Charge the
Unicorder via the Wall Charger |
Call Tech Support. The memory
card is full. Code 7. |
Memory card is
full, recording is not impossible. |
Initialize the
device to clear error. If the message
persists, please contact Support |
Lie on your back, look at the
ceiling and do not move. |
Startup command for Initializing the
study once the device has been turned on. |
Lie on your
back as instructed. Will hear
additional audio when okay to move around |
Initialization
successful. You can now go to sleep. |
Initialization
period has finished. |
The
patient can situate his/her desired posture and proceed to sleep. |
Tighten the nasal cannula. |
Alarm signaling
the nasal cannula has fallen off or that there is inadequate airflow
(pressure) causing the alarm to sound. |
Patient is
instructed to inspect the device and tighten the cannula back in place on the
device if necessary. |
The Unicorder has fallen off. |
Alarm signaling the device has fallen
off the head. |
Reposition the
Unicorder on head to resume testing. |
Adjust the Unicorder on your
forehead. |
Optical Signal
is disrupted. |
Adjust the
Unicorder on the forehead to clear error.
If the message persists, please contact the office that provided the
device. |